– 3 –
2.0 Testing Procedure
1. Place the instrument on the material
to be tested. The AD-100 can be used at
any angle, it does not need to be vertical.
NOTE: If the gauge is equipped with the
MemoryPointeroptionmakesurethe
pointer is returned to zero before starting.
2. Pressthefootofthegaugermly
against the specimen, but not so
rmlyastoimbedthefootintothe
surface of the material.
3. Maintainpressurefor2to3seconds.
The dial hand gives the reading in
durometer points.
IMPORTANT: After the initial reading has
been noted, continue maintaining pressure
forseveralmoreseconds.Creeporcold-owofthespecimen,ifpresent,
will be evident by receding action of the dial hand.
4. Repeat above steps for each use.
An example of a properly noted durometer reading:
“Durometer A 61, Creep 7 at 15 seconds, 73°F.”