5.0 OVERVIEW
01. ON/OFF Button Switches the instrument on/off
02. RESET Button For zero-setting after inserting the test sample
03. PEAK Button For selecting the display function, current measurement
result or peak value of the tractive force (drag indicator
function), and for calling up the peak value of the
measurement
04. MODE Button For selecting the desired measurement unit
05. Display
06. Outlet For AC Adapter
07. Outlet For data transfer
08. Terminal adapter with 12 slots for inserting the cable joint
09. Tractive force fixture for inserting the free cable end (with protection cap)
10. Selector switch for pull speed (access protected by screw connection)
5.1 Operating elements
– 16 –
10.0 TROUBLESHOOTING
No indication on display Plug in the AC-adapter and check the electrical
connections.
RS-232C communication Check the wiring of the connection cable and the
is disturbed compliance of wiring within the input terminals
of the PC-card.
Error Code Indications:
OV+ Overload of +120%. Reduce the load until the
instrument indicates a value within the admissible
measuring range.
OV- Overload of –120%. Reduce the load until the
instrument indicates a value within the admissible
measuring range.
OVP Overload +!! The load cell can be damaged. Remove the
excessive load immediately and check the instrument. If
the indicated measuring values are obviously incorrect,
the load cell must be replaced by the manufacturer.
OVM Overload -!! The load cell can be damaged. Remove the
excessive load immediately and check the instrument. If
the indicated measuring values are obviously incorrect,
the load cell must be replaced by the manufacturer.
ERR –3- Reading error of the E_proms. Switch off the instrument
and turn it on again. Should the error still exist, the
microprocessor may need to be replaced by the
manufacturer.
ERR –4- Reading error of the E_proms. Switch off the instrument
and turn it on again. Should the error still exist, the
microprocessor may need to be replaced by the
manufacturer.
Pull force fixture Remove the AC-adapter/charger, close the protection
does not work cover and wait for approx. 5 minutes. Plug in the
AC-adapter again and switch on the instrument. If the
display does not show any of the error codes above and
still does not work, please send the instrument back for
inspection to your dealer or directly to the
manufacturer.